Item Infomation


Title: Finite element analysis comparing WaveOne Gold and ProTaper Next endodontic file segments subjected to bending and torsional load
Authors: Ismail, Amira Galal
Participants: Zaazou, Mohamed Hussein Abdelfattah
Galal, Manar
Kamel, Nada Omar Mostafa
Nassar, Mohamed Abdulla
Issue Date: 2019
Series/Report no.: Bulletin of the National Research Centre, Volume 43 (2019), Article number: 194
Abstract: ProTaper Next® revealed higher flexibility than WaveOne Gold® when exposed to cantilever bending but showed higher stress accumulation than WOG. In terms of torsional resistance, PTN also revealed higher torsional resistance than WOG. The geometry of the instrument, thermomechanical treatment of the alloy, and its composition affect the mechanical behavior (bending and torsion) of nickel titanium rotary files. Hence, being aware of these behavioral differences, each clinician will be able to use the adequate file according to the clinical situation in addition to the manufacturer’s instructions.
URI: http://tailieuso.tlu.edu.vn/handle/DHTL/11104
Source: https://bnrc.springeropen.com/articles/10.1186/s42269-019-0215-6