Item Infomation
Title: | Finite element analysis comparing WaveOne Gold and ProTaper Next endodontic file segments subjected to bending and torsional load |
Authors: | Ismail, Amira Galal |
Participants: | Zaazou, Mohamed Hussein Abdelfattah Galal, Manar Kamel, Nada Omar Mostafa Nassar, Mohamed Abdulla |
Issue Date: | 2019 |
Series/Report no.: | Bulletin of the National Research Centre, Volume 43 (2019), Article number: 194 |
Abstract: | ProTaper Next® revealed higher flexibility than WaveOne Gold® when exposed to cantilever bending but showed higher stress accumulation than WOG. In terms of torsional resistance, PTN also revealed higher torsional resistance than WOG. The geometry of the instrument, thermomechanical treatment of the alloy, and its composition affect the mechanical behavior (bending and torsion) of nickel titanium rotary files. Hence, being aware of these behavioral differences, each clinician will be able to use the adequate file according to the clinical situation in addition to the manufacturer’s instructions. |
URI: | http://tailieuso.tlu.edu.vn/handle/DHTL/11104 |
Source: | https://bnrc.springeropen.com/articles/10.1186/s42269-019-0215-6 |
ISSN: | 2522-8307 |
Appears in Collections: | Tài liệu mở |
ABSTRACTS VIEWS
12
VIEWS & DOWNLOAD
0
Files in This Item:
There are no files associated with this item.
Bạn đọc là cán bộ, giáo viên, sinh viên của Trường Đại học Thuỷ Lợi cần đăng nhập để Xem trực tuyến/Tải về
Items in DSpace are protected by copyright, with all rights reserved, unless otherwise indicated.